NSK Ti-Max nanoSG20LS Manuel D'opération page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

CAUTION
• Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
• Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
• Immediately after use, the product should be cleaned, lubricated and sterilized. If blood remains on the external or internal surfaces it can
become clotted and cause rust.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
• If the sterilizer chamber temperature may exceed 135°C during the dry cycle then delete the dry cycle.
• Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
• Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
NOTICE
• NSK recommends Class B sterilizers as stated in EN13060.
13

Publicité

Table des Matières
loading

Table des Matières