Maintenance Schedule - RR Mechatronics Starrsed RS Mode D'emploi

Masquer les pouces Voir aussi pour Starrsed RS:
Table des Matières

Publicité

14.2 - Maintenance schedule

Maintenance Schedule Starrsed RS (Example)
Sample volume: 300 per day (5 working days)
or earlier after number of samples
Perform End-of-day wash
Clean outside aspiration needle
Check tubing/diluent syringe
Clean outside instrument
Clean Fill nozzle
Clean Liquid separator
Check sensors
Replace Fill nozzle O-ring
Replace bacterial air filter (HEPA)
Replace Disc filter (white) Only if internal waste
container is used
Replace Rinse tube assembly
Replace Saline tube assembly
Run Fill and clean
132
Daily
Weekly
Level 4
(monthly)
7.500 samples
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Level 3
Level 2
Level 1 (yearly)
(quarterly)
(halfyearly)
23.000 samples
46.000 samples
91.000 samples
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Annexe pour l'Starrsed RS
Parts
Total amount/
year
Per 91.000
samples
X
X
X
X
X
X
X
X
QWLV050004
X
QWLV040002
X
QWLV040001
X
ESRI090902
X
ESRI090903
X
MRN-074-2-FR Starrsed RS
12
12
12
12
12

Publicité

Table des Matières
loading