Data Recording; Other Technical Specifications; Operating Environment; Classification - Gima 35132 Manuel De L'utilisateur

Moniteurs de signes vitaux
Masquer les pouces Voir aussi pour 35132:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3. TEMP responding time: ≤150s for KRK sensor; ≤40s for YSI sensor

6.5 Data Recording

1. Sensitivity selection tolerance: ±5%
2. Recording speed: 25mm/s
3. Recording speed accuracy: ±10%
4. Hysteresis: ≤0.5mm
5. Frequency response: 0.5~40Hz for normal mode, 0.05~40Hz for enhanced mode.
6. Time constant: ≥0.3s for normal mode, ≥3.2s for enhanced mode.

6.6 Other Technical Specifications

1. AC power supply voltage: 100~240VAC
2. AC power frequency: 50/60 Hz
3. Battery specification: 11.1V/4400mAh (Li-ion Battery)

6.7 Operating Environment

Working Environment
Ambient temperature range: 5°C ~ 40°C
Relative humidity: 30 ~ 80%
Atmospheric pressure: 70kPa ~106kPa
Transport and Storage Environment
Ambient temperature range: -20°C ~ 60°C
Relative humidity: 10 ~ 95%
Atmospheric pressure: 50.0kPa ~107.4kPa

6.8 Classification

Safety standard
The type of protection against electric shock
The degree of protection against electric shock
Electro-Magnetic Compatibility

6.9 ECG Monitoring

1. Input signals range in amplitude: ± (0.5 mVp ~ 5 mVp)
2. Heart rate display range: 15 bpm ~ 350 bpm
3. Heart rate display accuracy: ± 1% or ± 2 bpm, whichever is greater.
4. Heart rate averaging: Averages the recent eight beats having RR intervals falling within the acceptable
5. Defibrillation recovery time: ≤10 sec
6. Alarm signal (for any alarm source) generation delay time: <1 sec
User Manual for Vital Signs Monitor
limits.
50
IEC 60601-1
Class I equipment.
Type BF, CF applied parts
Group I, Class A