Cleaning of the MAYFIELD Infinity XR2 Base Unit
Do Not Steam Sterilize! The carbon Fiber material and plastic components
may be damaged by heat.
Following these steps is recommended:
1. Remove base unit from operating table support rails.
2. Remove Swivel Adaptor, and Link Arm Assembly from base unit
3. The Infinity XR2 Base Unit should be thoroughly cleaned after each use. Scrub each
component with a soft brush and use only a pH-neutral detergent. Clean thoroughly to
remove any traces of blood and/or debris and to prevent such blood or debris from
interfering with function or movement. Rinse thoroughly with clean water.
4. Dry all components with a soft dry towel.
5. After components are totally dry, re-assemble the components. Unit is ready for next use.
Optional Automatic Wash Cycle:
Any deviation from this guideline could result in damage to equipment as well
as improper cleaning results. Parameters given below must be followed exactly
in order to assure proper cleaning and avoid damage to the equipment.
Disassemble the device as indicated in section Cleaning of the MAYFIELD Infinity XR2
Base Unit.
Rinse components under warm tap water prior to placing in washer.
Arrange device in a way to avoid contact between components.
Select instrument cycle and ensure proper programming.
WARNING!
CAUTION!