Operation

Disinfect contaminated surfaces of the curing light as
well as light probes and anti-glare cones before each
use. Furthermore, make sure that the stipulated light
intensity permits adequate polymerization. For that
purpose, check the light probe for contamination and
damage as well as the light intensity at regular intervals.
Selecting the curing time
The curing time can be individually set using the time
selection button. Users may choose between 10, 15, 20
and 30 seconds.
Observe the Instructions for Use of the material applied
when selecting the curing time. The curing recommenda-
tions for composite materials apply to all shades and, if
not mentioned otherwise in the Instructions for Use, to a
maximum layer thickness of 2 mm. Generally, these
re commendations apply to situations where the emission
window of the light probe is placed directly over the
Curing times
Restorative materials
Indirect restorations /
cementation materials
Adhesives
Temporary materials
Miscellaneous
1) Applies to a maximum layer thickness of 2 mm and provided that
the Instructions for Use of the respective material do not state any other
recommendation (might be the case e.g. with dentin shades)
2) Applies to a maximum layer thickness of 4 mm and provided that
the Instructions for Use of the respective material do not state any other
recommendation (might be the case e.g. with dentin shades)
Composites
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow
Tetric EvoCeram / Tetric EvoFlow
Tetric / Tetric Basic White
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric EvoFlow Bulk Fill
Tetric Basic White
Heliomolar / Heliomolar HB / Heliomolar Flow
All conventional composites
1)
Compomers
3)
Compoglass F / Compoglass Flow
Variolink Esthetic LC
/ Variolink Esthetic DC
4)
Multilink Automix
/ SpeedCEM Plus
5)
AdheSE / Adhese Universal / ExciTE F/ ExciTE F DSC
Heliobond / Syntac
Telio CS Link
Telio CS Inlay/Onlay
Telio Stains
Telio Add-On Flow
Helioseal / Helioseal F / Helioseal Clear
Heliosit Orthodontic
IPS Empress Direct Color
IPS Empress Direct Opaque
MultiCore Flow / Multicore HB
material to be polymerized. Increasing the distance
between the light source and the material will require
the curing time to be extended accordingly. For instance,
if the distance to the material is 9 mm, the effective light
output is reduced by approx. 50%. In this case, the
recommended curing time has to be doubled.
per mm ceramic: 10 seconds per surface
5)
per mm ceramic: 20 seconds per surface
5)
3) Applies to a maximum layer thickness of 3 mm
4) Applies to light-curing
5) Applies to dual-curing

Start button

Time selection
button
HIGH POWER
1,200 mW/cm
± 10%
2
10 seconds
10 seconds
15 seconds
20 seconds
10 seconds
10 seconds per surface
10 seconds
10 seconds
15 seconds
20 seconds
10 seconds
20 seconds
20 seconds
20 seconds