www.allice.de
Inrush protection:
IP20 per IEC 529.
Shock protection:
100g, 6ms, half-period, per IEC 68-2-27.
Drop height, all angles:
1 metre.
Vibration resistance:
10/55/10Hz, 0.15mm, per IEC 68-2-6.
Bumps:
40g, 6ms, 4,000 bumps, per IEC 68-2-29.

4.4. ELECTRIC SAFETY

Protection against electric shocks
Per IEC 61010-2-032, 600V, CAT III, pollution degree 2.
Electromagnetic compatibility
Emissions and immunity in an industrial setting compliant with IEC 61326-1.
Ip (Apeak)
Allice Messtechnik GmbH
Nondestructive peak current limit as a function of frequency
140
100
50
0
10
100
1 000
10 000
22
f (Hz)