VWR INCU-Line ILS4 Manuel D'instructions page 17

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Other-1 Controller on at all
times - "locked-up"
Other-2 Front panel displays
are all off
Other-3 Unit or wall fuse/circuit
breaker is blown
Other-4 Unit will not turn on
Other-5 Unit is smoking – Out
of box
Other–
Contamination in
6
chamber
OTHER
1/ Turn unit off and on to reset.
2/ If cannot change any condition on the front panel,
call VWR Customer Service.
Check for wire damage.
1/ Check wall power source.
2/ Compare current draw and compare to specs on
data plate.
3/ See what other loads are on the wall circuit.
1/ Check wall power source.
2/ Check fuse/circuit breaker on unit or in wall.
3/ See if unit is on, e.g., fan or heater, and just
controller is off.
4/ Check all wiring connections, especially around the
on/off switch.
This is not an uncommon occurrence when first
operating new units. Put unit under vent and run at
full power for one hour. – smoking is normal during
first cycle to temperature.
1/ See cleaning procedure in see section 8.0.
2/ Develop and follow standard operating procedure
for specific application; include definition of cleaning
technique and maintenance schedule.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Incu-line ils6444-0732444-0733

Table des Matières