The effectiveness of ventilation and alarms should be verified including after any ventilation or
alarm setting change, any change in circuit configuration, or after a change to co-therapy (eg,
nebulisation, oxygen flow).
The Astral device and AC Power Supply can get hot during operation. To prevent possible skin
damage do not leave the Astral device or AC Power Supply in direct contact with the patient for
extended periods of time.
The device can provide therapies typically associated with both ventilator-dependent and
non-dependent patients. The mode of ventilation, circuit type, and alarm strategies should be
chosen after a clinical evaluation of each patient's needs.
The device must not be used at an altitude above 3000m or outside the temperature range of
32–104°F (0–40°C). Using the device outside these conditions can affect device performance
which can result in patient injury or death.
A caution explains special measures for the safe and effective use of the device.
CAUTION
Repairs and servicing of the device should only be performed by an authorised ResMed service
representative.
The temperature of the airflow for breathing produced by the device can be as much as 6ºC
higher than the temperature of the room. Caution should be exercised if the room temperature
is warmer than 35ºC.
Do not expose the device to excessive force, dropping or shaking.
A note advises of special product features.
Notes:
For assistance and reporting of issues associated with the Astral device, contact your Health Care Provider or
authorised ResMed representative.