CPR

Before resuscitation can/may be performed on the sys-
tem, the system must be fully deflated.
For rapid deflation, pull down on the red tab printed with
„CPR" at the head end (see photo.). The tab is on the
same side as the hose system. The valves open and air
escapes from all of the cells.
To supply the system with air once again, close the CPR
valve again by pressing it firmly and securely fasten the
tab marked „CPR" once again. Incl. photo (photo 6).
6

Transport function

Remove the hose system from the unit and promptly
close it with the attached cap. Now the mattress is in
transport mode. To return to normal operation, remove
the cap on the connection hose and connect it back to
the unit and start the unit.
Note: No patient should be positioned on the system
when it is in transport mode!

Power failure

In the event of a power failure, set the system to trans-
port mode. See section on transport function.

Alarm function

Low pressure
The „Low Pressure" LED is lit whenever the unit detects
too little air in the cells.
An acoustic signal sounds at the same time. Once the set
pressure has been reached in the cells again, the display
goes out and the acoustic signal stops.
Important: During the filling phase, the „Low Pressure"
LED is lit; once the pressure in the air chambers has
reached the set user weight, the „Low Pressure" LED
goes out.
Drive Medical GmbH & Co. KG | As per: 13.07.2020 | Errors and alterations excepted

Cleaning, disinfection and care

Mattress overlay system
The mattress overlay system should not be saturated or
soaked*, however it can be cleaned by hand. To do this,
use a moist (not wet) cloth and a mild standard cleaning
and disinfectant solution.
* There is a risk of liquid getting into the cell chambers
and hoses.
Unit
• Before cleaning, disconnect the unit from the power
supply. During cleaning, please ensure that no fluids/
cleaning solutions come into contact with the plug.
• The unit should not be saturated or soaked.
• The unit may be cleaned by hand. To do this, use a
moist (not wet) cloth and a mild household cleaning
and disinfectant solution.
• Do not use any solutions which contain phenol or abra-
sive cleaners. This can corrode the surface of the unit.
• Allow the unit to dry thoroughly.
Perform a function test after the device is thoroughly dry.
Cover
• The cover can be cleaned by hand. To do this, use a
moist (not wet) cloth and a mild cleaning solution.
• You can detach the cover using the zipper and wash
it at a max. temperature of 60° using a household
detergent.
• Do not dry the mattress cover in the dryer.
Note!
If the alternating pressure system is machine-washed
and -disinfected, a validated and automated method ac-
cording to the Robert Koch Institute (RKI) and/or the As-
sociation for Applied Hygiene (VAH) is necessary.

Reuse/patient change

• In the case of reuse/patient change, the alternating
pressure system must be disinfected and cleaned
according to a validated method recognized by the
Robert Koch Institute.
• During processing, take note of the medical device law,
the „Guidelines for the Processing of Medical Devices"
as well as the „Guidelines for Hospital Hygiene and
Infection Prevention" of the Robert Koch Institute.
EN
15
Instructions for use DUO WAVE