Technical Information

Interconnections

Interconnections

The following paragraphs describe the Turbo-V 3K-G
interconnections.
J1 – Remote I/O
At power-on the controller goes into "Remote I/O" mode, even if it
was in "Serial" mode before power-down.
This connector carries all the input and output signals to remote
control the Turbo-V 3K-G.
It is a 15-pins D type connector; the available signals are detailed in
the table, the following paragraphs describe the signal characteristics
and use.
PIN N.
1
2
3
4
5
6
7
8
310/360
SIGNAL NAME
START/STOP (+)
START/STOP (-)
INTERLOCK (+)
INTERLOCK (-)
SPEED SETTING (+)
FAULT COMMOMN
SOFT START (+)
SET POINT
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01 (D)
INPUT/OUTPUT
IN
IN
IN
IN
IN
OUT
IN
OUT
ISOLATION
Optocoupled
Optocoupled
Optocoupled
Optocoupled
Optocoupled
Relay
Optocoupled
Relay