Figure 2
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The Turbo-V 3K-G must be connected to a primary pump (see
"Technical Information").
The Turbo-V 3K-G can be installed in any position. Fix the Turbo-V
3K-G in a stable position connecting the inlet flange of the
turbopump to a fixed counter-flange capable of withstanding a torque
of 33000 Nm around its axis.
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01 (D)
maximum pressure: 2 bar above atmospheric pressure
air temperature: from +5 °C to +35 °C
relative humidity: 0 – 95 % (non-condensing)