Télécharger Imprimer la page

Profibus Message Mapping - Agilent Technologies Turbo-V 3K-G Notice De Mode D'emploi

Publicité

Profibus Message Mapping

Tab. 12
Parameter buffer (17 bytes)
Offset
Size
Unit Res
Byte
Byte
0
1
°C
1
1
°C
2
1
%
3
1
°C
4
1
sccm
5
2
7
2
Hz/mA/
s
9
2
s
11
2
0.2*s
13
2
0.2*s
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01 (D)
Range
Description
15-35
Water Cooling Max Temperature (not used – insert 20)
1-5
Water Cooling Hysteresis (not used – insert 2)
1-99
Setpoint Hysteresis.
Set the hysteresis on set point output (see set point out
paragraph)
-
Heather Jacket Temperature Setting
Specific option, not available on standard models
-
Flowmeter Threshold
Specific option, not available on standard models
RESERVED
-
Setpoint Value
Set the threshold for the set point out
0-99999
Setpoint Mask Time
Set the time during the controller does not drive the set
point out
0-65535
Vent Valve Opening Delay
0 = inf.
Set the delay time between the stop command and vent
opening.
Valid only in "purge/vent auto mode" (Parameter Buffer
byte-15 B1=0); default = 15; 0 = infinite (the valve will
never open)
0-65535
Vent Valve Open Time
0=inf
Set how long the vent will remain open, then both purge
and vent will be closed. Valid only in "purge/vent auto
mode" (Parameter Buffer byte-15 B1=0); 0 = infinite
(the valve remain open till a start command)
Technical Information

Profibus Message Mapping

17
Win
Serial
158
159
105
113
115
-
102
103
126
147
337/360

Publicité

loading

Ce manuel est également adapté pour:

96988889698889