Télécharger Imprimer la page

Interconnections; Earth (Ground) Connection - Agilent Technologies TwisTorr 404 FS Notice De Mode D'emploi

Publicité

17
Technical Information

Interconnections

Interconnections

The following figure shows the TwisTorr 404 FS interconnections.
CONTROLLER
CONNECTION
Figure 13

Earth (Ground) Connection

It is required to connect the pump to protective earth with a M5
screw (marked thread in the pump envelope) using a green/yellow
wire of size AWG 16. Please ensure a reliable connection with using
an M5 screw and a snap ring or washer disc of the same metal as the
pump body to prevent corrosion. The screw must be fixed with a 2
Nm torque.
The impedance between the pump-body and the earth connection
point must be < 0.1 Ω.
288/326
TwisTorr 404 FS User Manual / 87-901-048-01 (A.00)
EARTH GROUND