Inlet flange
Foreline flange
Rotational speed
Start-up time
Recommended forepump
Operating position
Operating ambient temperature
Relative humidity of air
Bakeout temperature
Lubricant
Cooling requirements
Coolant water
Noise Pressure level (pump only)
Pwr supply:
Input Voltage:
Input Frequency:
Max oper. power (± 5%):
Installation category
Pollution degree
Storage temperature
Max operating altitude
Weight kg (lbs):
CFF 6"
ISO 100K
ISO 100F
Mean values based on a significative sample (Ar and N2 compression ratio estimated);
(■)
standard deviation per test: noise pressure level: ± 10%.
Estimated values
(♦)
TwisTorr 404 FS User Manual / 87-901-048-01 (A.00)
CFF 6"
ISO 100F
KF25 NW
Auto setting from 40800 RPM to 49500 RPM
< 5 minutes
mechanical: DS302
dry pump: IDP-10
Any
+5 °C to +35 °C
0 – 90% (not condensing)
80 °C at inlet flange (ISO) and 120 °C (CFF) max
Permanent lubrication
Max forced cooling air temperature = 35 °C
Max cooling water temperature = 25 °C
Minimum flow: 100 l/h
Temperature: +15 °C to +25 °C
Pressure: 2 to 4 bar
43 dB(A) at 1 meter
2
(provided by the controller)
54 Vac, 3 Phase
825 Hz max
200 W water cooling
200 W air cooling
II
2
-40 °C to +70 °C
3000 m
22.4 kg (49.4)
22.4 kg (49.4)
22.4 kg (49.4)
Technical Information