17
Technical Information
Controller Connection
Window Meanings
Tab. 15
N.
000
001
008
100
101
102
103
244/306
Read/ Write
Data Type
R/W
L
R/W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Description
Admitted Values
Start/Stop
Start = 1
(in remote/ Front
Stop = 0
mode the window
is a read only)
Low Speed
No = 0
Activation
Yes = 1
(default = 0)
Remote
Remote = 1
(default) or Serial
Serial = 0
configuration
(default = 1)
Soft Start
YES = 1
(write only
NO = 0
in Stop
condition)
R1 Set Point type
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 3)
R1 Set Point
(default = 867)
valve
(expressed in Hz, W
or s)
Set Point delay:
0 to 99999
time between the
(default = 0)
pump start and the
set point check
(seconds)
(I)