TECHNICAL INFORMATION

CONTROLLER SPECIFICATIONS

Input:
Voltage
Frequency
Power
Output:
Voltage
Frequency
Power
water cooling
air cooling
Operating temperature
Storage temperature
Fuse (mains)
P1 optoisolator input
J2 optoisolator output
J6 output (fan)
J7 output (forepump coil, etc.)
Radio interference suppression
Auxiliary connectors
P1
J2
J6
J7
J14
Interconnecting cables
Weight (both models)
50
±
100, 120, 220, 240 Vac
47 to 63 Hz
600 VA maximum
±
54 Vac nominal
10%
±
700 Hz
2%
420 W maximum
320 W
220 W
0°C to +40 °C
-20°C to +70°C
2 x T4A(slow blow) for 220 or 240 input Voltage
2 x T8A(slow blow) for 100 or 120 input Voltage
Minimum ON 3mA
24 Vdc, 60 mA
120Vac, 1A disregarding the mains
120 Vac, 1A
disregarding the mains
EN 55011 class A group 1
EN 61010-1
IEC1000-4-2,1000-4-3, 1000-4-4
External INPUT signals (pins)
OUTPUT signals (sockets)
Output fan Voltage
Valve, forepump socket
RS 232/422/485 connection (optional)
Mains cable (3-wire, 3-meter long)
Pump cable (6-wire, 5-meter long)
10.7 kg (24 lbs)
TURBO-V 550 CONTROLLER - 87-900-873-01 (A)
10%, 1-phase
Maximum 5mA