Error/Warning
Description
Connect a Probe
Probe disconnected
or connected
improperly
Software not
updated to most
current version
Problem with probe
Probe Not
Probe disconnected
Supported
or connected
improperly
Software not
updated to most
current version
Problem with probe
HQd meter does not
support IntelliCAL
probe
Bootloader
Software not
X.X.XX.XX error
updated to most
current version.
Solution
Tighten the locking nut on the probe
connector.
Disconnect the probe and then
connect the probe again
Update the meter software to the
most current version:
www.hach.com/
SoftwareDownloads
Connect a different IntelliCAL
probe to verify if problem is with
probe or meter
Tighten the locking nut on the probe
connector.
Disconnect the probe and then
connect the probe again.
Update the meter software to the
most current version:
www.hach.com/
SoftwareDownloads
Connect a different IntelliCAL
probe to the meter to verify if
problem is with the meter or the
probe.
Contact Technical Support.
Update the meter software to the
most current version:
www.hach.com/
SoftwareDownloads
Error/Warning
Description
0 days remaining
LDO or LBOD
message (For LDO
sensor cap used for
and LBOD only)
365 days
There are 0 days
remaining in the life
of the LDO sensor
cap.
Meter set to
incorrect date and
time
Software not
updated to most
current version
Meter not
Software error(s)
configured
Solution
Replace the LDO or LBOD sensor
®
cap and iButton
.
Replace the LDO sensor cap.
Calibration will be allowed.
However, the calibration icon and
question mark will appear on the
measurement screen even if the
calibration has passed.
1. Disconnect the probe from the
meter.
2. Remove the meter batteries.
3. Install the meter batteries
properly. Follow the polarity
makings.
4. Set correct date and time in the
meter.
5. Connect the probe and verify
that message has been
removed.
Update the meter software to the
most current version:
www.hach.com/
SoftwareDownloads
If the meter starts up correctly, back
up the Data Log and Method files.
Update the meter software to the
most current version:
www.hach.com/
SoftwareDownloads
English 15