1. Push the
key.
2. Select, create or delete an operator ID:
Option
Description
Current ID
Select an ID from a list. The current ID will be
associated with sample data until a different ID is
selected.
Create a New
Enter a name for a new operator ID (maximum 10
Operator ID
names can be entered).
Delete Operator ID
Delete an existing operator ID.
Data management
About stored data
The following types of data are stored in the data log:
• Sample measurements: stored automatically each time a sample is
measured in the Press to Read or Interval Mode. When the continuous
measurement mode is used, data is stored only when Store is selected.
• Calibrations: stored only when Store is selected at the end of a
calibration. Calibration data is also stored in the IntelliCAL (R) probe.
• Check standard measurements: stored automatically each time a check
standard is measured (in the Press to Read or Interval Mode).
When the data log becomes full (500 data points), the oldest data point is
deleted when a new data point is added. The entire data log can be deleted
to remove data that has already been sent to a printer or PC (
Delete Data Log). To prevent deletion of the data log by a user, use the
Security Options menu.
View stored data
The data log contains sample, calibration and check standard data. The
most recent data point in the data log is tagged as Data Point 001.
1. Push the
key.
2. Select View Data Log to view the stored data. The most recent data
point is shown. The top of the screen shows whether the data is from
a sample reading, a calibration or a check standard. Push the
to view the next most recent data point.
Option
Reading Log
Calibration Log
Check Standard
Log
View stored probe data
Make sure that a probe is connected to the meter. If two probes are
connected, select the appropriate probe when prompted.
1. To view the calibration data that is stored in a probe, push the
and select View Probe Data. The current calibration and calibration
history for the probe can be viewed.
Option
View Current
Calibration
key >
View Calibration
History
Print stored data
The meter must connect to AC power to start the USB connection. Make
sure that the connection to AC power is made before the meter is powered
on.
All data can be sent to a printer. The PD-24 Citizen Printer is compatible
with the HQd meters and is FCC Part 15B, Class B compliant with the HQ
meters. Other printers may not be compatible. Compatible printers should
Description
Reading Log—shows sample measurements including
the time, date, operator and sample ID. Select Details to
view the associated calibration data.
Calibration Log—shows calibration data. Select Details to
view additional information about the calibration.
Check Standard Log—shows check standard
measurements. Select Details to view the calibration data
that was associated with the measurement.
Description
The current calibration information shows the calibration
details for the most recent calibration. If the probe has not
been calibrated by the user, the factory calibration data is
shown.
The calibration history shows a list of the times when the
probe was calibrated. Select a date and time to view a
summary of the calibration data.
key
key
English 11