Dispenser Operations; System Care; Recommended Maintenance; Beginning The Day - DENTSPLY Professional Cavitron DualSelect Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11
Section 9: Operation (con'd)
9.2

Dispenser Operations

• Turn on the scaler/JET. Fluid flow from the DualSelect
Dispensing System is controlled by the water control on the
ultrasonic scaler, Cavitron
®
Systems.
• Turn the Selector Knob to the desired position: Bottle "A",
"Open • H
O", or Bottle "B". Selector knob can only be rotated
2
180° through the "Open • H
"A". (as shown to right)
• The scaler/JET connected to the DualSelect
receive fluid from the selected source.
• When switching between bottles "A" & "B", allow 30 seconds
for new fluid to exit handpiece.
• To use office water, set Selector Knob to "Open • H
position.
NOTE: When utilizing closed line option, no water will exit
the system on "Open • H
2
line is connected to the DualSelect
position must be selected to receive fluid.
Section 10: System Care
10.1

Recommended Maintenance

Beginning the Day

1. Place bottle with clean water in position "A."
2. Close Shield and set Selector Knob to position "A."
3. Operate ultrasonic scaler/JET at maximum water flow
for at least one minute, per the Centers for Disease
Control and Prevention (CDC) guidelines.
4. Set selector knob to "Open • H
5. Lift shield, remove bottle from position "A" and place
it in position "B." Repeat steps 2 & 3 in position "B."
6. Open shield and remove bottle from "B" position.
7. Place bottle with clean water/medicament in one or
both positions, close shield and set Selector Knob to
appropriate position.
8. Place a freshly sterilized insert/nozzle into the hand-
piece.
9. Unit is now ready for operation.

Between Patients

1. Remove ultrasonic/JET insert or air polishing nozzle
from the handpiece, clean and sterilize.
2. Place bottle with clean water in position "A." Close
Shield and set Selector Knob to position "A."
3. Operate ultrasonic scaler/JET at maximum water flow
for at least 30 seconds, as per the Centers for Disease
Control and Prevention (CDC) guidelines,
to flush all previously used fluids from the system.
ENGLISH • 7
JET Systems or PROPHY-JET
O" setting; "A" to "B" or "B" to
2
System can now
O"
2
0" position since no external water
System. The "A" or "B"
O" position.
2
®
4. Set Selector Knob to "Open•H
Shield. Remove bottle with clean water from position
"A" and place it in position "B." Repeat steps 2 & 3
in position "B." Set selector knob to "Open•H
position, open shield and remove bottle with clean
water from position "B."
5. Clean and disinfect all surfaces of the DualSelect
Dispensing System and attachments by applying an
approved non-immersion type disinfectant solution
carefully following the instructions provided by the
disinfectant solution manufacturer. To clean system,
generously spray disinfectant solution on a clean towel
and wipe entire unit. Discard used towel. To disinfect
system, generously spray disinfectant on a clean towel
and wipe entire unit. Allow disinfectant solution air
dry.
6. Place bottle with clean water/medicament in one or
both positions, close Shield and set Selector Knob to
appropriate position.
7. Place a freshly sterilized insert/nozzle into the
handpiece.
8. Unit is now ready for operation.
NOTE: Water based disinfectant solutions are preferred.
Some alcohol based solutions may be harmful and may
discolor plastic materials.
"A" to "B"
"B" to "A"
O" position and open
2
O"
2

Publicité

Table des Matières
loading

Table des Matières