Table I - Calibration Information
Gas
Combustible
CL2
CO
H2S
* HF
NO2
O2
SO2
Refrigerant
Rxxx
* Discontinued products: Only existing sensors will be supported.
Scale
0 - 102% LEL 15 min.
0 - 15.3 ppm
15 min.
0 - 255 ppm
N/A
0 - 51.0 ppm
15 min.
0 - 10.2 ppm
15 min.
0 - 10.2 ppm
15 min.
0 - 25.5%
N/A
0 -10.2 ppm
15 min.
0 - 1000 ppm 15 min
301EM User Manual
Warm-up
Flow Rate
500 ml/min
500 ml/min
500 ml/min
500 ml/min
500 ml/min
500 ml/min
500 ml/min
500 ml/min
100 ml/min
Programming and Calibration
Calibrating the Unit
1% Volume CH4 =
20% LEL CH4.
Nitrogen dioxide is
a surrogate gas for
HF
Replace the xxx by
the required
refrigerant
Notes
47