Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used
in such environments:
IMMUNITY TEST
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
IEC 60601
COMPLIANCE
LEVEL
TEST LEVEL
3 Vrms
150 kHz
to 80 MHz
3 V/m
3 V/m
80 MHz
to 2.5 GHz
ENGLISH
ELECTROMAGNETIC ENVIRONMENT
– GUIDANCE
Portable and mobile RF communications
equipment should be used no closer
to any part of the device, including ca-
bles, than the recommended separation
distance calculated from the equa-
tion applicable to the frequency of the
transmitter.
Recommended separation distance
√P
d = 1.2
80 MHz to 800 MHz
√P
d = 2.3
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmit-
ters, as determined by an electroma-
gnetic site survey,
a
the compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
should be less than