To avoid inaccurate results caused by electromagnetic interference between electrical and electronic
equipments, do not use the device near a mobile phone or microwave oven.
At least keep a maximum output power of 2 W yields and a distance 3.3m away from this equipment.

APPENDIX

Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used
in such environments:
IMMUNITY TEST
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
ENGLISH
IEC 60601
COMPLIANCE
TEST LEVEL
LEVEL
± 6 kV contact
± 8 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
ELECTROMAGNETIC
ENVIRONMENT – GUIDANCE
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthe-
tic material, the relative humidity should
be at least 30%.
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.