Setting Up; Display - Endress+Hauser mini-log b 507000 3W Manuel De Mise En Service

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6. Setting up

6.1 Display

The following symbols can be displayed during operation.
1
2
3
Ê Low battery display:
In the battery version measurement continues for approx. 30
days after the low battery display is activated.
In order to achieve continuous recording the data should be
read out and the battery changed as swiftly as possible.
(See 6.4 Changing battery). In the remote powered version a
power failure will be ccovered for about 2 hours. After a long
power failure please check time and date. The measurements
previously stored will not be effected.
Ë 7 digit numeric block:
The actual values are displayed as a numeric block.
Every 5 seconds the display scrolls as follows:
Analogue value -> digital value -> time
Attention: Dependent on the unit settings the inputs can be
scrolled in a different manner. For example should the digital
input be inactive then it would not be displayed. The display
then scrolls from the analogue value to time every 5 seconds.
Ì Limit markings:
The unit can monitor two internal set points.
These are defined using the software package deliverd with the
unit.
Should the unit recognise a set point infringement it displays
this using the appropriate symbol.
Example: The left symbol will be the low set point and the right
symbol an upper set point.
30

Publicité

Table des Matières
loading

Table des Matières