8. FUNCTION CHECK

9. TROUBLESHOOTING

Problem
No weight displayed
with a load on the scale.
0.00 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display "StOP"
appears.
The display "tEMP"
appears.
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview".
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from
page 101.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 101, you may not use the device.
Have the device repaired by seca service or
an authorized specialist dealer.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on
A load was already on the scale before it was
switched on.
- Remove the load from the scale
- Switch the scale off and back on again
The corresponding point has a fault.
- Inform the service department
Maximum load has been exceeded.
- Remove the load from the scale
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C
- Wait for around 15 minutes until the device
has adapted to the ambient temperature