4. Attach the Hook Leg of the Cartridge onto one of the Support Rods while holding the tubing between the Cartridge
legs and against the Tubing Retainer (see Figure 7D).
5. Swing the Cartridge with the tubing over the rotor and push down on the Tab (see Figure 5) until the Latch snaps
closed (see Figure 7E). Note: Depress the Cartridge as required in the location indicated in Figure 7E to facilitate
latching and unlatching.
6. Lightly pull the tubing at the outlet and inlet to remove the slack when using continuous tubing. No slack removal is
required when using a Microbore Tubing Set.
7. Adjust the occlusion setting (see OPERATION section). For a nominal setting with MASTERFLEX tubing, turn ad-
justment screw until inside edge of the wedge is aligned midway between #3 and #4 on the label.

(3) Cartridge Removal

Follow this procedure to unload the Cartridge.
1. Turn the Drive OFF.
2. Push down on latch side of Cartridge (see Figure 7E) and pull Latch Leg outward to unhook from rod. The Latch Re-
lease (see Figure 5) can be depressed to aid in this process as desired.
3. Pivot Cartridge upwards and unhook Hook Leg from rod and remove Cartridge.

(4) Bi-Directional Pumping

Fluids can be pumped into and back out of containers by reversing the drive. On drives that cannot be reversed, use lon-
ger tubing lengths and simply reverse the loaded Cartridge.
7A
7B
7D
FIGURE 7 TUBING AND CARTRIDGE LOADING
7
7C
7E