Hach Polymetron 9526 Manuel D'utilisation De Base page 3

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 64
Table of contents
on page 3
on page 5
Additional information
Additional information is available on the manufacturer's website.
Specifications are subject to change without notice.
Analyzer
Specification
Dimensions
Weight
Casing protection
Power supply
Sample flow rate
Sample tubing
Connections
Ambient temperature
Maximum temperature 100 °C (at atmospheric pressure)
Maximum pressure
Relative humidity
Precision
Measurement range
Display resolution
on page 13
Details
Height: 450 mm; Width: 250 mm; Depth: 460 mm
7 kg (15.4 lb)
IP 65 / NEMA4X
Standard version: 100-240 VAC 50/60 Hz
Low voltage version: 13-30 VAC 50/60 Hz, 18-42 VDC
Consumption: 25 VA
Measurement category: I (overvoltage less than 1,500 V)
20 liters/hour minimum
Sample inlet and outlet: Diameter 8 mm (or 5/16'') semi-rigid tubing. We recommend the
use of PE tubing if sample temperature is inferior to 70 °C, and PTFE if superior to 70 °C
Power supply: Use the connector provided in the drawer
Analog output: Use the recommended POLYMETRON cable
-20 to 60 °C (-4 to 140 °F)
10 bar at ambient temperature
10-90%
Conductivity: ± 2% of the displayed value
Temperature: ± 0.2 °C
Conductivity: 0.01 μS/cm to 200 μS/cm
Resistivity: 100 MΩ.cm to 5 kΩ.cm
Temperature: -20 to 200 °C (-4 to 392 °F)
0.001 μS/cm or 0.1 MΩ.cm
English 3

Publicité

Table des Matières
loading

Table des Matières