Hach Polymetron 9526 Manuel D'utilisation De Base page 17

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 64
Calibration
About sensor calibration
There are no calibration options available from the menus of the 9526 instrument. All calibrations are
performed from the controller of the sensor under test. For detailed information on these calibration
procedures refer to the associated manuals delivered with the sensor and controller under test.
After turning on the 9526 instrument and initiating sample circulation, wait for at least 30 minutes in order to allow
correct rinsing of the whole system. This also enables the temperature equilibration between the sample, the flow
chamber and the sensor.
After 30 minutes of sample circulation, compare the measurement value displayed on the controller
under test against the measurement value displayed on the 9526 instrument. If these values are
outside ± 5% of each other than a calibration is required. If these values are within ± 5% of each
other than a calibration is not necessary but can still be performed.
Before calibrating the sensor under test, make sure to perform a temperature calibration first.
Calibration process
All calibrations are made using the controller and sensor under test. Follow the instructions in the
associated controller and sensor user manuals.
The process can be different according to the Polymetron controller under test. Perform the
calibration process in the following sequence.
1. Temperature calibration
The following equipment is required for a temperature calibration:
• Pt100 simulator (< 0.1 °C) for a 2-point electrical calibration
• Certified precision thermometer (< 0.1 °C) if connected in-line
• None if mounted off-line as the 9526 is used as the reference
2-point electrical
calibration
Process calibration
2. Electrical calibration
9500 / 9125
3
R∞
and 200 kΩ
3. Conductivity calibration
9500 / 9125 / 8920
K calculation (see
K Calculation
on page 18). On the controller
under test, enter the K Cell value
computed by the Polymetron 9526
3
Cable disconnected from the sensor or sensor exposed to air
4
Use the precision resistance on the 9526
5
Adjust the displayed value of the transmitter under test to that of the 9526 value
N O T I C E
Polymetron controller model under test
9500
NO
YES
Polymetron controller model under test
9125 (< V1.12) / 8925 / Others
4
Polymetron controller model under test
9125 (< V1.12) / 8925
Process: R∞ and comparative
measurement with 9526
9125
YES (100 and 172 Ω)
YES
NO
5
Others
NO
YES
8920
3
R∞
Others
Process: Comparative
5
measurement with 9526
English 17

Publicité

Table des Matières
loading

Table des Matières