 At small voltage ranges, unsteady readings will appear before the test leads
make contact with the circuit. This is normal since the meter is highly sensitive.
When the test leads are connected to the circuit, the tru readings will be shown.
AC Voltage
range
2 V
20 V
200 V
600 V
 Max. input voltage: 250 VDC at 200 mV range, 600 VDC elsewhere.
 Input impedance: 10 MΩ.
 Frequency response: 200 Hz at 600 V range, 40-400 Hz elsewhere.
 Response: average (RMS of sine wave).
Notes
 At small voltage ranges, unsteady readings will appear before the test leads
make contact with the circuit. This is normal since the meter is highly sensitive.
When the test leads are connected to the circuit, the tru readings will be shown.
DC Current
range
2 mA
20 mA
200 mA
10 A
 Overload protection:
o mA ranges: resettable fuse F1, 400 mA/600 V (quick acting).
o 10 A range: F2, 10 A/600 V fuse (quick acting).
 Max. input current:
o mA jack (mA range): 200 mA.
o 10 A range: 10 A.
AC Current
range
20 mA
200 mA
10 A
 Overload protection:
o mA ranges: resettable fuse F1, 400 mA/600 V (quick acting).
o 10 A range: F2, 10 A/600 V fuse (quick acting).
 Max. input current:
o mA jack (mA range): 200 mA.
o 10 A range: 10 A.
 Frequency response: 40-400 Hz.
 Response: average (RMS of sine wave).
V. 01 – 21/01/2015
DVM894
resolution
1 mV
10 mV
100mV
1 V
resolution
1 µA
10 µA
0.1 mA
10 mA
resolution
10 µA
0.1mA
10 mA
accuracy
± (0.8 % of readings + 3 digits)
± (1.2 % of readings + 2 digits)
accuracy
± (0.8 % of readings + 1 digit)
± (1.5 % of readings + 1 digit)
± (2.0 % of readings + 5 digits)
accuracy
± (1.0 % of readings + 5 digits)
± (1.8 % of readings + 5 digits)
± (3.0 % of readings + 7 digits)