Chapter 5 Working with the AVS 470 Viscosity Measuring Unit
° C
100 ° F
SUS
[g/dl]
no
0,00 s
1.23 s
Fig. 16 Creation mode following Fig. 15
Input of the temperature value to document and
select its temperature scale (°C or °F), if the "abs"
or "rel" mode was selected - press the "Enter" key
to proceed to Fig. 17 or to Fig. 19 in "relative"
mode.
Fig. 17 Creation mode following Fig. 16
Selection of the viscometer type from the list in
chapter 9
Input of the constants or guideline constants
Input of the number of measurements
Input of the temperature pre-adaptation time
Input of the maverick test, 0 = no, number = yes
Selection criterion in ± n.nn%
Selection of HC correction, 0 = no, 1 = yes.
Press "Enter" to proceed to selection in Fig. 20
Fig. 18 Creation mode following Fig. 16
If Sayboldt calculation was selected in Fig. 15, this
is the point to enter the working temperature in
terms of °F; subsequently, you have to specify SUS
or SFS for the calculation in the selection field
below.Press "Enter" to proceed to Fig. 17
Fig. 19 Creation mode following Fig. 17
If "relative" was selected in Fig. 15:
Selection of the calculation types: 0 = no selection,
1 = selection. Please note: eta rel. includes eta rel.
and VN eta rel. and eta spec.
Input of the concentration, and selection of its
dimension; only one value is possible.
Press "Enter" to proceed to Fig. 17
Fig. 20 Creation mode, confirmation prompt
following Fig. 17
This prompt is issued for data integrity reasons to
avoid the automatic take-over of any erroneous
input or changes. Yes means: accept and store
data No means: discard changes
Press "Enter" to return to Fig. 4 or, in "Relative"
mode, to Fig. 21 if "yes" was selected.
Fig. 21 "Relative" creation mode
If "Relative mode" or "Blank value" was selected in Fig.
15, you will be prompted here to know whether the
blank value (t
) is to be entered manually, or whether it
0
is to be measured instead. The measurement will be
made using the parameters which were input in Fig. 17.
If manual input is specified you will proceed to Fig. 14,
Measurement, pressing "Enter" will take you to Fig. 22.
Fig. 22 "Relative" creation mode, measurement
result
As soon as the measurement of a blank value is com-
pleted, the result will be displayed here; by selecting
the "press Enter" field you can accept the result as t
subsequently, you will be taken to Fig. 4.
;