54
methode : absolute
ready!
start
create methode
delete methode
system parameter
ViscoPump parameter
Id: Sample XYZ
lot: 12075ADC
usr: Jonny Miller
temp equ. time
122 s
No. 1 out of 10
120 s
meas. 2
245.56 s
result corr.
average = 1234.56 s
stand. dev. = 0.001
correktion =
corr. ave. = 1233.22
Abs Visc =
123.322 mm ^2 / s
Chapter 5 Working with the AVS 470 Viscosity Measuring Unit
1.34 s
Fig. 4.
Display of the set method (e.g. absolute)
Start the set method
Create/change the set method
Delete the method
Change/set the system parameters
Change&/set the ViscoPump parameters Fig. 23
Fig. 5 Input after start
Id: The designation of a method; in the present
case the field is empty (after a reset, or yet unused)
or contains the last input
lot: The lot designation (as above)
usr.: User (as above)
No mandatory entry, press "Enter" to proceed
Fig. 6 after Fig. 5
If "pre. temp." (fig. 17) was set, this screen will
display the remaining time of the running
temperature pre-adaptation in terms of seconds.
Fig. 7 Measurement after temperature pre-
adaptation
After temperature pre-adaptation, the set number of
measurements (Fig. 17) will be preformed and
displayed here on a running basis in steps of 1 s.
Fig. 8 Measurement value at the end of a
measurement
Upon completion of a measurement, the respective
result will remain on the display until the next
measurement begins.
Fig. 9 Display of the measurement result
Example: Absolute measurement with
correction
The present example shows the result of an
absolute measurement in which the Hagenbach-
Couette correction was applied. The average
shown corresponds to the (exemplary!) amount,
reduced by the number of correction seconds which
are also shown.