NAMUR commands
IN_NAME
IN_PV_X
X=1;2;3;4;;
IN_SOFTWARE
IN_SP_X
X=1;2;3;4;12;42
52;54;
IN_TYPE
OUT_NAME name
OUT_SP_12@n
OUT_SP_42@n
OUT_SP_X n
X=1;2;4;52; 54;
OUT_WD1@m
OUT_WD2@m
RESET
RMP_CONT_X
X=1;4
RMP_IN_X
X=1;4
RMP_IN_X_y
X=1;4
RMP_LOOP_SET_X
X=1;4
RMP_LOOP_RESET_X X=1;4
RMP_OUT_X_y n
X=1;4
hh:mm:ss
RMP_PAUSE_X
X=1;4
RMP_RESET_X
RMP_START_X
X=1;4
RMP_STOP_X
X=1;4
START_X
X=1;2;4;5;7
Function
Title request
Current value reading
Software ID number, date, version request
Set target value reading
Lab device identification request
Output of identification name. (Max. 6 characters; default: IKAHBR)
Setting WD safety limit temperature with set value echo
Setting WD safety limit speed with set value echo
Setting of target value to n
Watchdog mode 1: if event WD1 should occur, the heating and stirring func-
tions are switched off and ER 2 is displayed. Set watchdog time to m (20
- 1,500) seconds, with watchdog time echo. This command launches the
watchdog function and must be transmitted within the set watchdog time.
Watchdog mode 2: if event WD2 should occur, the speed target value is
changed to the WD safety speed limit and the temperature target value is
changed to the WD safety temperature limit value. The warning WD is dis-
played. The WD2 event can be reset with OUT_WD2@0 - this also stops
the watchdog function. Set watchdog time to m (20 - 1,500) seconds, with
watchdog time echo. This command launches the watchdog function and
must be transmitted within the set watchdog time.
Switches off the device function.
Continuaton of ramp function (After prior RMP_PAUSE_X).
Reading the real segment number of ramp. With ramp not started: 0
Reading the accumulated value and the ramp segment duration of ramp seg-
ment y.
Ending of ramp loop.
To work off the ramps in one loop.
Setting the accumulated value (n) and the ramp segment duration (hh:mm:ss)
for ramp segment y.
Stopping the ramp function. Freezing of real rated value and real ramp seg-
ment time.
Switching off ramp functions and deleting of all set ramp segments
Starting the ramp function, beginning with ramp function No.1 (Only possible
after prior START_X. After RMP_STOP_X START_X is not necessary.)
Switching off ramp function. Rated value = 0. (Ramp is maintained, that
means, ramp can be restarted with RMP_STARTX_X).
Starting the instrument's (remote) function.