Zero point adjustment (OFFSET)
12 
1x 
short 
13 
 1x long 
The instrument must be adjusted until the nominal value is reached! 
 
If „ERR 07" appears in the display the OFFSET value is out of the acceptable 
range (  Technical Data). The instrument must be returned to the 
manufacturer for checking. 
Changing the unit of measurement mm / inch (UNIT)
You can select between „mm" and „inch".  
 
The gage must be switched on. The SETUP – Menu can be activated in each program. 
 
Activating
 1x long 
„UNIT" will appear in the display 
 3x 
 
short
Activating UNIT – Menu 
 1x long 
The last used measuring unit will appear in the display / Factory setting: mm 
Changing the measuring unit 
 short 
For confirming and saving 
 1x long 
 
Afterwards you will return to the SETUP ‐ Menu. 
Quitting the SETUP – Menu 
3x 
 
short
For confirming and saving. 
The instrument returns to the last program used 
 1x long 
 
After changing the unit the tolerance limit must be checked again, as the tolerance limit will 
be rounded automatically to the new unit and could be incorrect. 
 
14 
Quitting the SETUP – Menu 
For confirming and saving. 
The instrument returns to the last program used 