Kroeplin C Série Mode D'emploi page 31

Palpeurs extérieurs électroniques
Masquer les pouces Voir aussi pour C Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

 
Program: MIN / MAX / HOLD 
Important: 
The measuring axis 
must be in vertical 
position to the 
measuring  
 
surface!
 
 
Important: 
The measuring axis must 
be in vertical position to 
!
the measuring surface
 
Important: 
The fixed flat contact 
must be square against 
the part to be measured! 
Hold gage steady around 
2s!  
 
Determination of the minimum measuring value via pivoting 
the gage (optimal measurement) 
Measured value   
too high  
 
Determination of the maximum measuring value for measurement of 
narrow grooves via lateral displacement of the gage, where the gage 
cannot be pivoted vertically. 
Measured value   
too small 
 
For checking round parts it is advisable to use measuring gages with 
chisel contacts (  Technical Data). 
Measurement of flat material (e.g. foam material) with flat measuring 
contacts. Take note of the measuring force! 
The movable flat contact aligns optimally so a pivoting or moving of the 
gage is not necessary. 
 
Fixed flat 
contact 
 
Measured value                      Measured value 
 
correct   
 
Measured value   
 
correct   
Movable flat 
contact 
 
                  too high 
Measured value 
 
           too small 
9

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C110C110tC110sC1r10C1r10sC220 ... Afficher tout

Table des Matières