Absolute - Mode / Relative - Mode - Kroeplin C Série Mode D'emploi

Palpeurs extérieurs électroniques
Masquer les pouces Voir aussi pour C Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

 
Absolute – Mode / Relative – Mode
If the „Absolute – Mode" is activated measurements will be made with reference to the 
zero point. 
(This mode will be started during the initial use or after factory setting) 
 
If the „Relative – Mode " is activated measurements will be made with reference to the 
reference value (gage block) which was selected before. This mode is used for 
comparison measurement. 
 
 
Switching between „ Absolute– / Relative – Mode " can be made in MIN / MAX / HOLD 
and TOL. 
 
 
Switch‐over from „Absolute – Mode" into „Relative – Mode ": 
 
The instrument must be switched on. 
Finding the reference value via block gage block 
 
(example: 10mm) 
Zero point adjustment according to the reference 
value 
1x short 
Symbol (4.1) for „Relative‐Mode" appears in the display 
 
footer, ...... 
............. as well as the reference value (example: 
 
10mm). 
 
 
Switch‐over from „Relative – Mode" into „Absolute – Mode": 
 
The instrument must be switched on. 
Example: 
 
Setting the zero point (10,00mm) 
The reference value is activated again 
1x short 
 
If „ERR 08" appears in the display the reference point in „Relative ‐ Mode" is out of the 
indicating range. 
 
 
 
 

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C110C110tC110sC1r10C1r10sC220 ... Afficher tout

Table des Matières