Connecting The Scale - Sartorius PMA.Vision VIS1X Notice D'installation

Balances pour peintures pour atmosphères explosibles de la zone 1
Masquer les pouces Voir aussi pour PMA.Vision VIS1X:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 32
3.4.1 AC Adapter Assembly
Using the wrong mains adapter may cause
a fatal electric shock and damage the
equipment.
Never plug the mains adapter into the socket
when it is disconnected from the AC adapter
(danger of electrical shock).
t
Use the right mains adapter for your mains power supply:
Mains adapter set YAK01
Bag
Region/Country
a) transparent
Europe/EU (except United Kingdom)
b) blue
USA
c) yellow
United Kingdom
Mains adapter set YAK02
d) red
Australia
e) turquoise
South Africa
f) white
Argentina
g) pink
Brazil
Mains adapter set YAK03
h) light brown
China
i) black
India
j) green
Korea
t
6
Installation Instructions VIS1X
Push (1) and slide (2) the mains
adapter required for your power
supply into the opening of the
AC adapter module.
When doing this, the mains
adapter needs to lock into
position.
Removing / Replacing the Mains
Adapter
t
Power Connection/Safety Precautions
− Only use original Sartorius AC adapters.
The AC adapter has an IP rating of IP40 in accordance
with N 60529/IEC 60529.
− Make sure that the voltage rating printed on this unit
matches the voltage at the place of installation.
− If the stated supply voltage or the plug design of the AC
adapter does not comply with your country's standard,
please inform your nearest Sartorius representative.
− The power must be connected in accordance with the
regulations applicable in your country.

3.4.2 Connecting the Scale

Connecting the Scale
t
t
t
Connecting the Grounding Cable
This explosion-protected system should be set up according
to commonly accepted technical standards. The applicable
national electrical code and safety regulations for your
particular country must be observed.
Before commissioning the scale, a check must be carried
out by or under the supervision of a qualified electrician
to ensure that the system is in good working order.
Check whether or not the competent authorities (e.g.
industrial supervisory board) need to be informed. It is also
necessary to carry out inspections of the system during
operation.
Unlock (1) and then remove the
mains adapter (2).
Insert the link cable plug into
the socket on the back of the
display.
Lay the link cable through the
cable holders on the back of the
scale.
Connect the link cable to the
ex-link converter.