9 Specifications

9.1 General Data

Specification
Scale, ID code: (explosion protection) II
2G Ex ib IIB T4 Gb as per EC Type
Examination Certificate no.
FM15ATEX0032X
Power supply
Input supply voltage
Power consumption
Other data
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment
Operational capability
Storage and shipping
Relative humidity
Ex-Link Converter Interface Connection
Electromagnetic Compatibility
Interference resistance
Transient emissions
Available Application Programs
AC adapter YPS07-USB
USB power plug (5 V/1500 mA)
Primary
Secondary
Other data
Ex-Link Converter YCO16-Z
ID code (explosion protection)
Other data
14
Installation Instructions VIS1X
Unit
Value
Only via USB interface or Sartorius AC adapter YPS07-USB
V
+5.0
DC
W
5.1
IP40 in accordance with EN 60529/IEC 60529
For indoor use only
°C
Guaranteed between +5 and +40
°C
-10 to +60
%
15% to 80% for temperatures up to 30°C non-condensing,
decreasing linearly to 50% relative humidity at 40°C
Ethernet
As per EN 61326-1/IEC 61326-1 Electrical equipment for measurement,
control and laboratory use – EMC requirements – Part 1: General
requirements
Basic requirements
Class B
Suitable for use in residential areas and areas that are connected to
a low voltage network that also supplies residential buildings.
Recalculation, factor calculation, formula
Type FW7721M (manufacturer description)
100–240 V~, -10%/+10%, 50–60 Hz, 100 mA
5 V
, ±5%, 1500 mA (max.)
DC
Protection class II
IP40 in accordance with EN 60529/IEC 60529
II (2)G [Ex ib Gb]IIB as per EC Type Examination Certificate no.
FM15ATEX0034X
IP40 in accordance with EN 60529/IEC 60529