Figure 1- 1: Dynamic And Offset Limitations - Tektronix P6245 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Configuration
To set the probe offset, follow these steps:
1. Set the oscilloscope coupling to GND.
2. Use the vertical position control to set a zero reference level on
the oscilloscope display.
3. Set the oscilloscope coupling to DC and 5 V/div.
4. Attach the probe to the circuit.
5. Adjust the probe offset to bring the trace to the oscilloscope zero
reference.
6. Change the volts/division setting to the desired range, adjusting
the offset to keep the trace on the zero reference level.
NOTE. The P6245 has a ±10 V offset range. The linear operating
range is ±8 V. See Figure 1- - 1. Also, see page 2- - 1 for more
information.
If cursors are used on a TEKPROBE oscilloscope, the zero reference
will be at the probe offset voltage.
Nonoperating Range (+15 V Maximum Non-destructive Input Voltage )
+15 V
+10 V
+8 V
0 V
- - 8 V
- - 10 V
- - 15 V
Nonoperating Range (- - 15 V Maximum Non-destructive Input Voltage )

Figure 1- 1: Dynamic and Offset Limitations

1- 8
Maximum AC Signal Amplitude
+10 V
Maximum Offset Range
- - 10 V
P6245 Instruction Manual

Publicité

Table des Matières
loading

Table des Matières