Fault/Remedy

Fault
No function
The scale appear to be displaying
false weighing results
"Err" is shown on the display
"Lo" is shown on the display
"ErrL" is shown on the display
"ErrH" is shown on the display
"C" is shown on the display
Body Analyser Scale BA 1300-1304
Possible Cause
‚ Battery empty or inserted the wrong way
around?
‚ Is the scale standing on a firm, even
surface?
‚ Is a device that can cause electromagnetic
disturbances located in the vicinity of the
scale?
‚ Remove the battery from the scale and wait
a moment. Reinsert the battery. The data
stored up to this point will be retained.
‚ The maximum capacity of the scale (180 kg)
has been exceeded.
‚ The battery is empty (see "Changing the
battery").
‚ The fat proportion % is too low: Below 4%
nothing can be displayed
‚ The fat proportion % is too high: Above
60% nothing can be displayed.
‚ The weighing/measurement process was
faulty: The scale are not standing stably or
you are not standing still on the weighing
surface during weighing.