20.1

Centrifugation with preselected time

Set the centrifugation parameters or call up a program or a program linkage (see the chapters on "Enter
centrifugation parameters", "Calling up programs" or "Program linkage").
Press the
key. The LED in the
START
Once the time has run up or if the centrifugation run is canceled by pressing the
takes place with the selected run-down parameter. The run-down parameter is displayed e.g. 9. The righthand
LED in the
OPEN / STOP
switches off and OPEN OEFFNEN is displayed. The right-hand LED in the
off, the left LED in the
OPEN / STOP
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber (only for
centrifuge with cooling), and the remaining time are displayed.
20.2

Continuous operation

Set the minutes, seconds and hours to "0" or start a continual run program (see chapter "Enter centrifugation
parameter" or " Calling up programs").
Press the
key. The LED in the
START
The time count starts from 00:00.
Press the
key to end the centrifugation run. Run- down is carried out with the selected run- down
OPEN / STOP
parameters. The run-down parameter is displayed e.g. 9. The right-hand LED in the
the rotor has come to a standstill, the LED in the
right-hand LED in the
OPEN / STOP
continues blinking until the lid is opened.
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber (only for
centrifuge with cooling), and the elapsed time are displayed.
20.3

Short-time centrifugation

A short-time centrifugation is not possible if program linkages are being used.
Keep the
key pressed. The LED in the
START
subsequently lit. The time count starts from 00:00.
Release the
key again to end the centrifugation run. Run- down is carried out with the selected run- down
START
parameters. The run-down parameter is displayed e.g. 9. The right-hand LED in the
the rotor has come to a standstill, the LED in the
right-hand LED in the
OPEN / STOP
continues blinking until the lid is opened.
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber (only for
centrifuge with cooling), and the elapsed time are displayed.
21

Change the settings during the centrifugation run

It is not possible to change centrifugation parameters during the centrifugation run if program linkages are
used or if a program lock is set. The functions of the various program locks are described in the chapter
"Setting program locks".
The runtime, speed, relative centrifugal force (RCF), start-up and run-out parameters as well as temperature (only in
centrifuges with cooling) can be changed during the centrifugation run.
Change the value of the desired parameter (see chapter, "Enter centrifugation parameter").
The changed setting is saved at program place "0" (see chapter "Automatic temporary storage"). The original
program is not overwritten. The program place number is displayed in parentheses ().This means that the
centrifugation data in the display no longer corresponds to the centrifugation data from the program place that
has been saved.
AB5650DEENFRIT
key blinks until the rotor has been imported, it is subsequently lit.
START
key is lit. Once the rotor has come to a standstill, the LED in the
key starts blinking and continues blinking until the lid is opened.
key blinks until the rotor has been imported, it is subsequently lit.
START
START
key also switches off, the left LED in the
START
START
key also switches off, the left LED in the
Rev. 16 / 04.2020
OPEN / STOP
OPEN / STOP
key switches off and OPEN OEFFNEN is displayed. The
OPEN / STOP
key blinks until the rotor has been imported, it is
key switches off and OPEN OEFFNEN is displayed. The
OPEN / STOP
EN
key, run-down
key
START
key also switches
key is lit. Once
OPEN / STOP
key starts blinking and
key is lit. Once
OPEN / STOP
key starts blinking and
77/220