Invacare® Kirilos / Atlante

7 Maintenance

7.1 Maintenance and Safety Inspection

WARNING!
Risk of Falling
Maintenance MUST be performed only by
qualified personnel.
Improper assembly may cause injury or damage.
– Regular maintenance of patient lifts and
accessories is necessary to assure proper
operation.
– DO NOT overtighten the mounting hardware.
This will damage the mounting bracket.
Service Interval
At normal daily operation, a service check-up should
take place every year, according to the Safety Inspection
Checklist. When performing annual or regular maintenance,
all parts designed to carry load must be, as a minimum,
tested with maximum load. All safety features must be
checked according to EN ISO 10535:2006 Annex B.
General Maintenance
Regular cleaning will reveal loose or worn parts,
enhance smooth operation and extend the life
expectancy of the lift.
Follow the maintenance procedures described in
this manual to keep your patient lift in continuous
service.
34
The Invacare Patient Lift is designed to provide a maximum
of safe, efficient and satisfactory service with minimum care
and maintenance.
It is important to inspect all stressed parts, such as slings,
hanger bar and any pivot points for signs of wear, cracking,
fraying, deformation or deterioration. All parts of the
Invacare patient lift are made of the best grades of steel,
but metal to metal contact will wear after considerable use.
Replace any defective parts immediately and ensure that the
lift is not used until repairs are made. Refer to the Safety
Inspection Checklist for specific information regarding wear
items.
There is no adjustment or maintenance of either the casters
or brakes, other than cleaning, lubrication and checking
axle and swivel bolts for tightness. Remove all debris, etc.
from the wheel and swivel bearings. If any parts are worn,
replace these parts immediately.
If you question the safety of any part of the lift, contact
your Dealer or Invacare representative immediately and
advise him/her of your problem.