Option
Standard Adjust
Dilution Factor
Standard
Additions
Select the reading mode
Select Single Wavelength, Multi Wavelength or Time Course from the SETTINGS>Photometer
Functions menu.
Select Single Wavelength options
1. Push SETTINGS>Photometer Functions>Single Wavelength>Options>Advanced Options.
2. Select the applicable options.
Option
λ
Concentration
Resolution
Save to User Programs Save the selected parameter as an unique user program. Select the name, unit,
Select Multi Wavelength options
1. Push SETTINGS>Photometer Functions>Multi Wavelength>Options>Advanced
Options>λ1–λ4.
2. Select the applicable options.
Option
Description
Wavelength
Select two or more wavelengths. Defaults: 400, 500, 700 or 800 nm.
Abs Formula Select the applicable absorbance formulas for the calculation of the multi-wavelength reading.
The formula specifies the wavelength and the coefficients.
Factors
Select the multiplication factors for converting absorbance values into concentration values.
Select Time Course options
1. Push SETTINGS>Photometer Functions>Time Course>Options.
2. Select the applicable options.
Option
Total Minutes
Interval Seconds Enter the interval in seconds for the Time Course. Range: 10 to 600. Default: 30 seconds
Wavelength
Data Recall
Description
Change stored calibration. Complete a test on a known standard at a concentration near
the top of the test range. Use this function to adjust the result to align the standard
concentration.
Enter a corrective dilution factor for specific characteristics. The entered number prompt
will be multiplied by the result for the adjustment. For example, if the sample has been
diluted by a factor of 2, set the dilution factor to on and enter 2.
Note: When a dilution is in effect, the dilution icon is shown on the display.
Examine the reading accuracy. Refer to the method procedure for more information.
Description
Select a wavelength between 340 to 800 nm. Default: 560 nm
Select the applicable unit and enter the factor. Defaults: mg/L and 1.000
Select the resolution with the necessary number of decimal places. Default: 0.01
wavelength, resolution, chemical formulas 1–4, calibration formula, upper and
lower limit and the timer 1–4.
Description
Enter the total time in minutes for the Time Course. Range: 1 to 60. Default: 10 minutes
Select a wavelength between 340 and 800 nm. Default: 560 nm
Show the saved Time Course data.
English 13