Guidance and manufacturer's declaration - electromagnetic immunity

The Skin IQ 1000 MCM is intended for use in the electromagnetic environment specified below. The customer or user of the
Skin IQ 1000 MCM should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
Recommended separation distances between portable and mobile RF communications equipment
and the Skin IQ 1000 MCM
The Skin IQ 1000 MCM is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Skin IQ 1000 MCM can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Skin
IQ 1000 MCM as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
150 kHz to 80 MHz
not applicable
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separate distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1, At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from surfaces, objects and people.
12
IEC 60601
Test Level
±8 kV Contact
±15 kV Air
±2 kV for power
supply lines
±1 kV for
input/output lines
1 kV line(s) to Line(s)
2 kV line(s) to Earth
5% half cycle
40% for 5 cycle
70% for 25 cycle
5% for 5 seconds
30 A/M
NOTE: U, is the a.c. mains voltage prior to application of the test level.
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
3.5
d=[ — ]
E1
N/A
0.12
N/A
0.37
N/A
N/A
N/A
Compliance Level
±8 kV Contact
±15 kV Air
±1 kV lines to ±2 kV
for power supply lines
1 kV line(s) to Line(s)
2 kV line(s) to Earth
5% half cycle
40% 5 cycles
70% 25 cycles
5% for 5 seconds
30 A/M
meters
P
1.2
3.7
12
Electromagnetic Environment Guidance
In accordance with IEC 60601-1-2,
if the floors are covered with synthetic
material, the relative humidity should be
at least 30%
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
800 MHz to 2.5 GHz
7
P
d=[ — ]
E1
0.23
0.74
2.3
7.4
23