arjo Skin IQ 1000 Microclimate Manager Mode D'emploi page 17

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 52
Guidance and manufacturer's declaration - electromagnetic immunity
The Skin IQ 1000 MCM is intended for use in an electromagnetic environment specified below. The customer or user of the
Skin IQ 1000 MCM should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Conducted RF
IEC 61000-4-6
150K - 80 MHz
Radiated RF
IEC 61000-4-3
80 MHz - 2.5 GHz
NOTE 1, At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the Skin IQ 1000 MCM is used exceeds the applicable RF compliance
level above, the Skin IQ 1000 MCM should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the Skin IQ 1000 MCM.
b) Over the frequency range 150kHz, field strengths should be less than 3 V/m.
Compliance Level
Level
3Vrms
3 Vrms
150K - 80 MHz
3 Vrms
3 Vrms
80 MHz - 2.5 GHz
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Skin IQ
1000 MCM, including cables, than the recommended
separation distance calculated from the equation
application to the frequency of the transmitter.
Recommended Separation Distance
Battery Operated Device
3.5
P 80 MHz to 800 MHz
d=[ — ]
E1
7
P 800 MHz to 2.5 GHz
d=[ — ]
E1
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (see
note a) should be less than the compliance level in
each frequency range (see note b).
Interference may occur in the vicinity of equipment
marked with the following symbol: