12. Maintenance
Check that everything is clean and working before using.
Recommended at least once per year
Check that
General
The Walker feels firm/free from play.
The Walker does not rattle during
manoeuvres.
The Walker is level and all of its cas-
tors are in contact with the floor.
The Walker is not dirty.
Armrest
Armrests are intact and clean.
Armrest widening is functional.
Handle
Grips are not damaged/dirty.
Handle adjustment is functional.
Battery
Check that the charging cable is
connected to the control box.
Check that the hand control is con-
nected to the control box.
Check that connections are made
according to the wiring diagram.
Check that the battery, actuators and
control box are not loose.
Height adjustment
Raising and lowering are functional.
The Walker feels stable at maximum
settings.
There is no play between the vertical
frame and bottom frame.
Height adjustment clamps lock.
The electric actuator fitting must be
free from play.
Height adjustment via hand control is
functional. The electric actuator must
run
smoothly at a constant speed.
Frame parts
There is no mechanical damage.
There are no scratches.
End plugs/lower frame fitted
Castors/brakes
Castors roll easily/tread not damaged.
The castors are firmly fastened to the
lower frame.
Castor brakes functional on all
castors.
Handbrake functional.
DHG can provide upon request neccessary technical information in order to assist maintenace and repair of the
device.
I F U
Action
Clean with lukewarm soap solution or alcohol-based cleaning agent (no
petroleum products).
Clean; replace if damaged.
Fit lock screws/plates as required.
Clean with lukewarm soap solution (no petroleum products).
Replace handle grips.
Fit lock screws/plates as required.
Fit new charging cable; this must always be connected to the control
box (IP65 protection).
Connect or fit new hand control as required.
Refer to the wiring diagram in the control box wiring section.
Tighten; replace with new fasteners as required.
Fit new guide sleeves (in frame) or end plugs (in chrome tubes) as
required.
Tighten all lock bolts.
Replace clamps.
Tighten the attachment fitting concerned; replace bolts/lock nuts as
necessary.
Check that the battery/hand control/ actuator are connected in accord-
ance as in the control box wiring section. Charge the battery. Contact
customer services.
Refer to the wiring diagram in the control box wiring section.
If damage is present, contact DHG customer services.
Touch up as necessary.
Fit new end plugs.
Clean or replace castors. NOTE: Castors are always fixed to the lower
frame with thread locker or lock.
nuts. Do not dismantle the castors; replace the whole castor assembly
instead.
Tighten the castor bolt and fit locking nut or use thread locker (depends
on model).
Replace castor assembly.
Adjust the brake or fit new castors.
9