10. Battery Charging

Plug needs to connect to 230V wall socket. Charging starts automatically.
o LED / battery flashes green.
The battery is of the LiOn type and insensitive to how long / how often you charge the battery. Full charge on
discharged battery takes about 24 hours.
We recommend that the battery plug be removed from the control box if the product will be out of use for an
extended period (more than one week) or if the battery is switched off by means of the ON/OFF switch. This
eliminates the risk of disabling the battery as the control box always uses a small quantity of electricity even
when the Walker is not in use.
11. LEDs on Control Box and Battery
Control box – green light when power is available from battery/mains socket.
Battery:
The LED flashes green during charging.
ON/OFF: To use the product, press ON.
The green lamp will begin to flash.
Use the OFF button (hold down for 5 s) to switch off the battery if the product
will not be used for an extended period.
The LED will shows a constant green light when the battery is fully charged.
When the battery is in normal state/in operation, the LED flashes green at
REGULAR intervals, namely 0.5 s ON, 4 s OFF.
When the battery requires charging (remaining charge <25%) the LED will flash
orange/red at REGULAR intervals, namely 0.5 s ON, 4 s OFF.
NOTE: The battery has an in-built device that automatically switches it off when
remaining charge is <20%. The LED is then extinguished. The battery must now be
fully charged before the electronics can be used again.