Guidance and Manufacturer's Declaration – Emissions All Equipment and Systems
This device is intended for use in the electromagnetic environment specified below. The customer or user of this device
should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
Class B
RF Emissions
Radiated and Conducted
CISPR 11
Emissions
Harmonics
Class A
IEC 61000-3-2
Flicker
Complies
IEC 61000-3-3
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC 61000-4-2
Electrical Fast
Transient/burst
±2kV on AC Mains
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
>95% Dip for
0.5 Cycle
Voltage dips, short
60% Dip for
interruptions and
5 Cycles
voltage variations
on power supply
30% Dip for
input lines
25 Cycles
IEC 61000-4-11
>95% Dip for
5 Seconds
Electromagnetic Enforcement – Guidance
This device uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
This device is suitable for use in all establishments including
domestic, and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Compliance Level
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
±6kV contact
tile. If floors are synthetic, the relative humidity
±8kV air
should be at least 30%.
Mains power quality should be that of a typical
±2kV on AC Mains
commercial or hospital environment.
±1kV Differential
Mains power quality should be that of a typical
±2kV Common
commercial or hospital environment.
>95% Dip for
Mains power quality should be that of a typical
0.5 Cycle
commercial or hospital environment. If the user
60% Dip for
of this device requires continued operation
5 Cycles
during power mains interruptions, it is
30% Dip for
recommended that the device be powered
25 Cycles
from an uninterruptible power supply or
>95% Dip for
battery.
5 Seconds
en - 15