Specification
Temperature response
Insulator
Connector
Controller
Specification
Component description
Operating temperature
Storage temperature
1
Enclosure
Power requirements
Altitude requirements
Pollution
degree/Installation
category
Outputs
Relays
Dimensions
Weight
EMC requirements
Korean registration
CE compliance
1
Units that have the Underwriters Laboratories (UL) certification are intended for indoor use only
and do not have a NEMA 4X/IP66 rating.
4 English
Details
< 30 seconds
PSU
Glass polyester (IP65)
Details
Microprocessor-controlled and menu-driven controller that operates the sensor and
displays measured values.
-20 to 60 ºC (-4 to 140 ºF); 95% relative humidity, non-condensing with sensor load
<7 W; -20 to 50 ºC (-4 to 104 ºF) with sensor load <28 W
-20 to 70 ºC (-4 to 158 ºF); 95% relative humidity, non-condensing
NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 W
sensor/network module load, 100 VA with 28 W sensor/network module load
(optional Modbus, RS232/RS485, Profibus DPV1 or HART network connection).
24 VDC powered controller: 24 VDC—15%, + 20%; Power 15 W with 7 W
sensor/network module load, 40 W with 28 W sensor/network module load (optional
Modbus, RS232/RS485, Profibus DPV1 or HART network connection).
Standard 2000 m (6562 ft) ASL (Above Sea Level)
Polution Degree 2; Installation Category II
Two analog (0-20 mA or 4-20 mA) outputs. Each analog output can be assigned to
represent a measured parameter such as pH, temperature, flow or calculated values.
Optional module supplies three additional analog outputs (5 total).
Four SPDT, user-configured contacts, rated 250 VAC, 5 Amp resistive maximum for
the AC powered controller and 24 VDC, 5A resistive maximum for the DC powered
controller. Relays are designed for connection to AC Mains circuits (i.e., whenever
the controller is operated with 115 - 240 VAC power) or DC circuits (i.e., whenever
the controller is operated with 24 VDC power).
½ DIN—144 x 144 x 180.9 mm (5.7 x 5.7 x 7.12 in.)
1.7 kg (3.75 lb)
EN61326-1: EMC Directive
Note: This is a Class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
User Guidance for EMC Class A Equipment
업무용을 위한 EMC 등급 A 장치에 대한
사용자 지침
사용자안내문
A 급 기기 ( 업무용 방송통신기자재 )
이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시
기 바라며 , 가정외의 지역에서 사용하는 것을 목적으로 합니다.
EN61010-1: LVD Directive