DENTAURUM POLYCLAV Mode D'emploi page 16

Autoclave de polymérisation des résines autopolymérisables
Masquer les pouces Voir aussi pour POLYCLAV:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

4.3. Notes on hardening cold-curing polymers in the Polyclav
pressure vessel
Before loading the Polyclav
been saturated with water. If the plaster model is dry, the formation of bubbles in
the acrylic will be unavoidable.
Fill the Polyclav
pressure vessel with water up to approx. 3 – 5 cm below the brim
®
(temperature approx. 40 – 46 °C / 104 – 114 °F).
Use a compressor or compressed air (max. 8.0 bar) to achieve an operating pressure
of 1.8 – 2.2 bar. The manometer monitors the pressure level.
The Polyclav
pressure vessel remains under pressure until the acrylic is completely
®
hardened.
Observe the Instructions for use for the acrylic used.
The Polyclav
pressure vessel can be attached via the pressure hose
®
(accessory, REF 070-015-00) to any compressor in a dental office or laboratory,
provided the compressor operates with more than 3 bar.
5. Maintenance
We recommend you clean the inside of the lid once a week to remove acrylic and
limescale. Any standard descaling agent can be used to remove limescale.
Pointed or corrosive cleaning materials or sharp, pointed instruments should not be
used in order to avoid damaging the surface.
Please note that sharp-edged objects such as fixators, articulators etc. may damage
the surface.
Heating the Polyclav
components and gaskets.
Please maintain the lid gasket clean and free of grease. If necessary, sprinkle with
talcum powder.
16
pressure vessel, make sure the plaster models have
®
pressure vessel with no water inside may damage mechanical
®
®

Publicité

Table des Matières
loading

Table des Matières