The display shows the first parameter, "P1 Lin".
IN CASE THE ZONE OF USE IS DIFFERENT FROM THE CALIBRATION ZONE ONE SHOULD:
1)
Set the gravity acceleration value for the ZONE OF CALIBRATION, see P5 step.
2)
Carry out the calibration as following described (P1 or P2 step).
3)
Set the gravity acceleration value for the ZONE OF USER, see P5 step.
P1 Lin
CALIBRATION WITH DEFINED SAMPLE WEIGHTS (*)
By pressing ZERO it's possible to access the calibration with defined sample weights. One is asked to enter the
password.To enter the menu press the M+, PCS, and % keys, one after the other.
-
The display shows "UNLOAD";
-
Empty the platter, after Stable indicator on the weight will be acquired automatically;
-
The display shows "L
-
Put the first sample weight, defined in the table below, on the scale and the stability indicator is on the weight will
be acquired automatically;
-
The display shows "L
-
Put the second sample weight, defined in the table below, on the scale and when the stability indicator is on the
weight will be acquired automatically;
-
The display shows "L
-
Put the thirdsample weight, defined in the table below, on the scale and when the stability indicator is on the
weight will be acquired automatically;
-
The indicator returns to weighing.
CALIBRATION WEIGHTS
MODEL
1° SAMPLE WEIGHT 50g
2° SAMPLE WEIGHT 100g
3° SAMPLE WEIGHT 150g
P2 CAL CALIBRATION WITH FREE SAMPLE WEIGHT (*)
By pressing ZERO it's possible to access the calibration with free sample weight.
-
The display shows "UNLOAD";
-
Empty the platter, after Stable indicator on the weight will be acquired automatically;
-
Set the value of the mass weight by using the MODE key to clear the value, the TARE key to increase the value
of the digit, the PCS key to go on the previous digit and the % key to go on the following digit.
-
Confirm the entered value with ZERO; the display shows "LOAD";
-
Put the mass on platter, after stable indicator on the weight will be acquired automatically.
-
The indicator returns to weighing.
P3 Cnt A/D CONVERTER POINTS
By pressing ZERO the instrument shows the A/D converter points relative to the weight on the scale.
Press PRINT/M+ to exit the step.
P4 AZ
ZERO TRACKING (*)
By pressing ZERO one sets the divisions' number for the zero tracking, in other words, scale's thermal drift
compensation parameter; the set value corresponds to the number of divisions which is cleared.
Press TARE to select one of the suggested divisions: oFF, 0.5d, 1d, 2d, 4d.
P5 GrA GRAVITY COEFFICIENT (*)
Trough this step one access the acceleration value setting of instrument calibration and use zone.
Manual entry of the g value: Set the value by using the MODE key to clear the value, the TARE key to increase the
value of the digit, the PCS key to go on the previous digit and the % key to go on the following digit; one must modify
all the 6 digits of the gravitational acceleration. Confirm the entered value with ZERO or press PRINT/M+ to exit
the step.
P6 CAP SCALE CAPACITY (*)
By pressing ZERO one selects the total capacity of the scale.
XXX";
XXX";
XXX";
HLD150
HLD300
100g
200g
300g
HLD600
HLD1500
200g
500g
400g
1000g
600g
1500g
12
HLD3000
HLD6000
1000g
2000g
2000g
4000g
3000g
6000g