Service Information; Static Discharge Precautions - Stryker Prime 1105 Manuel D'entretien

STATIC DISCHARGE PRECAUTIONS

English
The electronic circuits in the stretcher are completely protected from static electricity damage only while the stretcher
is assembled . It is extremely important that all service personnel always use adequate static protection when servicing
the electronic systems of the stretcher . Whenever you are touching wires, you should be using static protection.
Static Protection Equipment
The necessary equipment for proper static protection is:
• 1 static wrist strap; 3M part number 2214 or equivalent
• 1 grounding plug; 3M part number 61038 or equivalent
• 1 test lead with a banana plug on one end and an alligator clip on the other; Smith part number N132B699 or
equivalent
CAUTION
All electronic service parts will be shipped in static shielding bags . Do not open the bags until you have completed steps
2 and 3 of the following procedure . Do not place unprotected circuit boards on the floor . All circuit boards returned to
Stryker should be shipped in the static shielding bags the new boards were shipped in .
Static Protection Procedure
1 .
Turn off the main power switch at the foot end of the stretcher .
2 .
Unplug the power cord from the power outlet . To unplug, grasp the mold near the outlet and pull the cord in a
direction parallel to the floor (not at an angle) .
3 .
Insert the grounding plug into a properly grounded hospital grade power source . Plug the banana plug of the test
lead into the receptacle on the grounding plug . Connect the alligator clip on the other end of the test lead to the
ground chain of the stretcher
4 .
Place the static control wrist strap on your wrist . Connect the alligator clip at the other end of the wrist strap cord .
Return To Table of Contents